Sign In | Join Free | My chinacsw.com |
|
Brand Name : Zhaoli
Model Number : DC2430PD01
Certification : CE, ROHS, REACH
Place of Origin : China
MOQ : 2000
Payment Terms : L/C, T/T, Western Union
Supply Ability : 10,000pcs per day
Delivery Time : 25-45 days
Packaging Details : cartons + pallets
Name : DC air pump
Rated voltage : 6.0VDC
Bearing : sleeve bearing
Packing : Customized packing
Printing : Customized printing
Application : portable nebulizer/ blood pressure
DC Air Pump Diaphragm Pump For Portable Nebulizer Blood Pressure 6.0VDC
Pump type: diaphragm pump
Application: This model is mainly applied in blood pressure or any other atomizer use.
Specification:
Rated voltage | 6.0VDC |
Operating voltage | 4.0 - 7.2 VDC |
Rated current | The highest current is less than 380mA while pressurizing with DC6.0V from 0 to 300mmHg. |
Inflation time | It takes Less than 10 seconds to pressurize with DC6.0V from 0 to 300 mmHg in a 500cc tank. |
Leakage | Pressurizing from 0 till 300±30mmHg in a 500CC tank, then holding for 30 seconds. After that, measuring the dropped pressure value, which shall be less than 2.0mmHg/min. |
Max pressure | The maximum pressure value shall be more than 400 mmHg while pressurizing with DC 6.0 V in a 500CC tank from 0 till dead-head pressure value. |
Restart voltage | Fill pressure until 200mmHg pause,and then add DC 3.0V voltage pump should restart. |
Dielectric strength | 600VAC/1S leakage current <=5mA |
Temperature rise | <60K |
Drawing:
![]() |
Blood Pressure DC Air Pump Motor 6.0V 300mmHg For Portable Nebulizer Images |